| Specifications |
| Ranges: | 101E: 0 – 50 ppb to 0 – 10 ppm H2S or up to 20 ppm SO2; 102E: 0 – 50 ppb to 0 – 10 ppm TRS or 20 ppm SO2 (Note: For ranges higher than 0-10 ppm, please consult Teledyne API) |
| Units: | ppb, ppm, mg/m3 |
| Zero noise: | 0.2 ppb (RMS) |
| Zero drift: | < 0.5 ppb / 24 hrs, 1 ppb / 7 days |
| Span drift: | < 0.5 % full scale / 24 hrs, 1% full scale / 7 days |
| Lower Detectable Limit (LDL): | < 0.4 ppb |
| Rise and Fall Time: | < 120 seconds |
| Sample flow rate: | 650 cc/min ±10 % |
| Linearity: | 1% of full scale |
| Converter Temperature: | 101E: 315°C; 102E: 850°C |
| Dimensions (HxWxD): | 7” (178 mm) x 17” (432 mm) x 23.5” (597 mm) for analyzer. Model 102E requires same size for converter |
| Weight: | 101E: 41 lbs (18.3 kg); 102E: 53 lbs (24 kg) |
| Power: | 100v, 50/60 Hz (3.25A); 115V, 60 Hz (3.0A): 220-240V, 50/60 Hz (2.5A) |
| Analog Outputs: | 10V, 5V, 1V, 0.1V, selectable |
| Current Output: | Optional 4-20 mA isolated outputs (select up to 3 channels) |
| RS-232: | Standard, DB-9 connector |
| Status (Digital): | 8 outputs, 6 inputs (opto-isolated), 4 alarm outputs (optional) |
| Approvals: | CE |